Sunday, March 6, 2016

Preliminary specification for the open DC multimeter

I think to make 2 version of the open DC multi-meter :
- first one (LA stands for low accuracy) with a basic DC accuracy ~ 100 ppm which will be the test bench for the project
- second one (HA stands for high accuracy) with a basic DC accuracy ~ 20 ppm



OSHW DMM Specs
Ranges Accuracy per year Resolution Other characteristics and features
DC Voltage ranges
 
1V
10V
 
~ 100 ppm (LA)
~ 20 ppm (HA)
1uV
 
Input resistance > 1 GOhms
Noise floor < 10 uV
Auto range
DC Current ranges
 
5A
1mA


To be estimated later
 
1mA
1uA


 
Burden voltage :
< 0.02 mV/mA for the 5A range
< 0.02 mV/uA for the 1mA range

Auto range
Resistance ranges
 
10KOhms
1MOhms


To be estimated later
 
0.08 Ohms
8 Ohms


 
1.25 mA current source
12.5 uA current source
2/4 wire measurement
Auto range

No comments: