I think to make 2 version of the open DC multi-meter :
- first one (LA stands for low accuracy) with a basic DC accuracy ~ 100 ppm which will be the test bench for the project
- second one (HA stands for high accuracy) with a basic DC accuracy ~ 20 ppm
- first one (LA stands for low accuracy) with a basic DC accuracy ~ 100 ppm which will be the test bench for the project
- second one (HA stands for high accuracy) with a basic DC accuracy ~ 20 ppm
Ranges | Accuracy per year | Resolution | Other characteristics and features |
---|---|---|---|
DC Voltage ranges
|
|||
1V 10V |
~ 100 ppm (LA) ~ 20 ppm (HA) |
1uV | Input resistance > 1 GOhms Noise floor < 10 uV Auto range |
DC Current ranges
|
|||
5A 1mA |
To be estimated later | 1mA 1uA |
Burden voltage : < 0.02 mV/mA for the 5A range < 0.02 mV/uA for the 1mA range Auto range |
Resistance ranges
|
|||
10KOhms 1MOhms |
To be estimated later | 0.08 Ohms 8 Ohms |
1.25 mA current source 12.5 uA current source 2/4 wire measurement Auto range |